Task #963602
openHtal HIGA San Martin - La Plata - Interfaz cs2500 (Sysmex)
0%
Description
Se realizaron pruebas el día 11/05/2023 con la siguiente interfaz
CA500
90009718M 041 116 042 969 043 095 051 279
Muestra: 90009718
041, 042, 043,051 son los código de enlace
116, 969, 095, 279 son los resultados
Los resultados no se guardaron en Nobilis
Se adjunta logs.
Files
Updated by Juan Poceiro 5 months ago
- File 10-05-23 Log SYSMEX CS2500 COM.txt 10-05-23 Log SYSMEX CS2500 COM.txt added
- File 10-05-23 Log SYSMEX CS2500.txt 10-05-23 Log SYSMEX CS2500.txt added
- File 09-05-23 Log SYSMEX CS2500 COM.txt 09-05-23 Log SYSMEX CS2500 COM.txt added
- File 09-05-23 Log SYSMEX CS2500.txt 09-05-23 Log SYSMEX CS2500.txt added
Adjunto logs de la interfaz cs2500 de otro sistema por si les útil.
Updated by Nestor Matellan 3 months ago
Fernando Peralta wrote:
Consultamos si se realizó la prueba solicitada con:
XN1000
XNL
XT2000i
XS1000_ASTM
en modo HQ y cualquiera de los tipos de conexiones soportados (SERIAL O TCP)
En el caso de requerir LICENCIAS podemos realizarlas, con el fin de la prueba necesaria.
Quedamos a la espera de novedades, gracias
Updated by Juan Poceiro 3 months ago
El equipo se probo con la interfaz de CA500, que fue la que sugirió Pablo para realizar las pruebas.
Las pruebas se realizaron con conexión serial, en modo host query.
Los trazas del 11/05, son los logs de las pruebas realizadas con esa interfaz.
También se adjuntan logs del día 09/05 y 10/05 de otro sistema en donde esta operativa y funcionando la interfaz con el equipo.
Nota: Tener en cuenta que de tener que ir a realizar alguna prueba, este cliente esta a 100 Kilómetros, así que seria conveniente ese día de contar con desarrollo para realizar las pruebas en conjunto, y en lo posible cerrar la interfaz ese día.
Updated by Nestor Matellan 3 months ago
Solicitamos nuevamente se prueben especificamente con los analizadores que se detallan en el comentario inicial. Los logs con Nobilis que subieron no son ASTM, y según este manual este analizador es ASTM. Segun el manual es compatible con los modelos mencionados.
Updated by Juan Poceiro 3 months ago
Se coordino con Brenda Horrach , bioquímica que maneja el equipo, para realizar pruebas el día 11/07/2023.
Se realizaran pruebas con las interfaces solicitadas por desarrollo.
Updated by Juan Poceiro 3 months ago
El día 11/07/2023, se realizaron pruebas en modo host query con las siguientes interfaces
XN1000
XNL
XT2000i
XS1000_ASTM
Se adjunta los logs de transmisión de las distintas interfaces
Updated by Juan Poceiro 3 months ago
- File XN1000-20230711_163223-Pantalla.log XN1000-20230711_163223-Pantalla.log added
- File XN1000-20230711_163724-Pantalla.log XN1000-20230711_163724-Pantalla.log added
- File XN1000-20230711_163727-Channel.log XN1000-20230711_163727-Channel.log added
- File XN1000-20230711_163727-Interface.log XN1000-20230711_163727-Interface.log added
- File XN1000-20230711_163727-Pantalla.log XN1000-20230711_163727-Pantalla.log added
Interfaz XN1000
Updated by Juan Poceiro 3 months ago
- File XN-L-20230711_163435-Pantalla.log XN-L-20230711_163435-Pantalla.log added
- File XN-L-20230711_163455-Channel.log XN-L-20230711_163455-Channel.log added
- File XN-L-20230711_163455-Interface.log XN-L-20230711_163455-Interface.log added
- File XN-L-20230711_163455-Pantalla.log XN-L-20230711_163455-Pantalla.log added
Interfaz XNL
Updated by Juan Poceiro 3 months ago
- File XT2000i-20230711_164424-Pantalla.log XT2000i-20230711_164424-Pantalla.log added
- File XT2000i-20230711_164427-Channel.log XT2000i-20230711_164427-Channel.log added
- File XT2000i-20230711_164427-Interface.log XT2000i-20230711_164427-Interface.log added
- File XT2000i-20230711_164427-Pantalla.log XT2000i-20230711_164427-Pantalla.log added
Interfaz XT2000i
Updated by Juan Poceiro 3 months ago
- File XS1000IASTM-20230711_164040-Pantalla.log XS1000IASTM-20230711_164040-Pantalla.log added
- File XS1000IASTM-20230711_164042-Channel.log XS1000IASTM-20230711_164042-Channel.log added
- File XS1000IASTM-20230711_164042-Interface.log XS1000IASTM-20230711_164042-Interface.log added
- File XS1000IASTM-20230711_164042-Pantalla.log XS1000IASTM-20230711_164042-Pantalla.log added
Interfaz XS1000_ASTM
Updated by Juan Poceiro 3 months ago
Con las cuatro interfaz el equipo leyó el código de barra, realizo la consulta a las interfaces, pero nunca proceso las muestras.
Updated by Juan Poceiro 3 months ago
- File XT2000i-20230711_164835-Pantalla.log XT2000i-20230711_164835-Pantalla.log added
- File XT2000i-20230711_164838-Channel.log XT2000i-20230711_164838-Channel.log added
- File XT2000i-20230711_164838-Interface.log XT2000i-20230711_164838-Interface.log added
- File XT2000i-20230711_164838-Pantalla.log XT2000i-20230711_164838-Pantalla.log added
Se adjunta también logs de envió de resultados desde el equipo a la interfaz XT2000i
Updated by Juan Poceiro 3 months ago
- File 10-05-23 Log SYSMEX CS2500 COM.txt 10-05-23 Log SYSMEX CS2500 COM.txt added
- File 10-05-23 Log SYSMEX CS2500.txt 10-05-23 Log SYSMEX CS2500.txt added
- File 09-05-23 Log SYSMEX CS2500 COM.txt 09-05-23 Log SYSMEX CS2500 COM.txt added
- File 09-05-23 Log SYSMEX CS2500.txt 09-05-23 Log SYSMEX CS2500.txt added
Se adjunta nuevamente logs de la interfaz cs2500 de otro sistema que esta funcionando, por si les sirve para comparar con los logs generados en la pruebas con el Nobilis
Updated by Cristian Lobertti 3 months ago
Buenas tardes.
Adjunto archivo LIS892 Sysmex CS 2500 18072023.rar de prueba interfaz Sysmex CS2500.
Quedamos a la espera de resultados de las pruebas con el Jar enviado.
Updated by Juan Poceiro 2 months ago
- File CS2500-20230727_100946-Pantalla.log CS2500-20230727_100946-Pantalla.log added
- File CS2500-20230727_100951-Channel.log CS2500-20230727_100951-Channel.log added
- File CS2500-20230727_100951-Interface.log CS2500-20230727_100951-Interface.log added
- File CS2500-20230727_100951-Pantalla.log CS2500-20230727_100951-Pantalla.log added
El día 27/07/2023, se probo con el nuevo Jar enviado.
Se realizo el circuito completo, leyó etiqueta, realizo el host query en forma correcta, el equipo proceso y envió los resultados a la interfaz.
Pero los resultados no se guardaron en el sistema.
Adjunto log de la prueba.
Updated by Fernando Peralta 2 months ago
Gracias Gabriel.
Estuve observando los logs me llama la atención que hay varias alertas de No se encontró paridad correspondiente para la determinación de la muestra recibida: 041 042 043 051 061 062 y 511
Te consulto: Se elimino el analizador y volvió a agregar (pantalla analizadores administración) y configurar luego del reemplazo del ASTM. en la ultima prueba de logs enviados?
Updated by Juan Poceiro 2 months ago
El Viernes 28/07/2023, se realizaron pruebas en forma remota junto a Fernando Peralta, y Brenda que estaba en forma presencial.
Después de varias pruebas, con el nuevo Jar, se comprobó que los código de paridad en la interfaz, tanto el de envió como el de recepción, tiene que ser el mismo, y que toma el campo numérico para la paridad.
Nos confirmo Brenda, que el equipo lee las etiquetas, se realiza correctamente el host query y que llegan bien los resultados al Nobilis.
Quedo funcionando correctamente la interfaz.
Se adjunta configuración de las paridades de la interfaz .